Introduction: ^

Angstrom Advanced ADX-2700 θ-θ Powder X-ray Diffraction Instrument is multi-function diffractometer with exceptional analysis speed, reliability and reproducibility. The ADX2700 is a diffraction instrument designed for the challenges of modern materials research. ADX2700 can analyze powders, liquids, thin films, nanomaterials and many other different materials. The ADX2700 can be used for many different applications: Academic, Pharmaceuticals, Chemical & Petrochemical, Material Research, Thin Film Metrology, Nano technology, Food & Cosmetics, Forensics, Mining & Minerals, Metals, Plastics & Polymers, etc.

Features: ^

angstrom advanced adx2700 x-ray diffraction instrument featuresComputed tomography, High-resolution X-ray diffraction, High throughput screening, In-plane diffraction, Crystallite size and micro-strain analysis, Micro-diffraction, Non-ambient diffraction, Pair distribution function analysis, Phase identification, Phase quantification, Reflectivity analysis, Residual stress analysis, Crystallography, Texture analysis, Transmission, Thin film analysis.
angstrom advanced adx2700 x-ray diffraction instrument featuresADX-DWZ Combination of Eulerian cradle for stress and texture investigations, Thin film and Quantity Analysis attachment with control and analysis software with alignment-free feature.
ADCX sample changer is compact and rugged. Integrated spinning improves particle statistics in polycrystalline sample measurements. Fully automatic alignment. Programmable.
angstrom advanced adx2700 x-ray diffraction instrument featuresAngstrom Advanced AHTK 1000 high temperature attachment
Automated variable temperature stage for X-ray diffraction measurements of materials at elevated temperatures (room temperature-1200°C). The stage may be operated in vacuum. The sample is heated radiantly for reduced heat gradients within the sample. Automated z translation within the stage assures precise sample positioning even in the presence of thermal expansion of the sample.
angstrom advanced adx2700 x-ray diffraction instrument featuresAngstrom Advanced ALTK-450 Variable temperature attachment
Automated variable temperature stage for X-ray diffraction measurements of crystal structure (-193°C-450°C). The stage can be operated under liquid nitrogen cooling conditions.

Software: ^

General diffraction data processing by Angstrom Advanced: automatic peak search, manual peak search, integral intensity, separation of Kα1,α2, background remove, pattern smoothing and magnifying, multiple plot, three-dimensional plot and simulation of XRD pattern.
  • Qualitative Analysis: The data processing software has the search and match function on the base of whole profile and diffraction angle. The whole profile matching procedure employs the designed mode to do the qualitative analysis by reducing the search range from major, minor, to micro phase without indicating the diffraction angle. The diffraction angle matching procedure is based on the peaks position and intensity and usually used for the qualitative analysis of the data with large angle error.
  • Quantitative Analysis: After the phase composition is determined, the content of each phase could be calculated with the help of RIR or/and the Rietveld refinement (Quantitative Analysis without criterion)
  • Plot and Export: The data processing software is operated within the Windows interface. The preparing exported pattern could be labeled, zoomed in, zoomed out and also copied and pasted.
  • Phase identification, structure analysis, Thin film analysis, stress investigation, Texture analysis are all available

Parts and Specifications: ^

X-ray Generator Control mode 1kV/step, 1mA/step controlled by PC
Rated output power 4 kW
Tube voltage 10-60 kV 1kV continuously adjustable
Tube current 5-80 mA continuously adjustable
X-ray tube Cu, Fe, Co, Cr, Mo et al (2 kW)
Focus dimension: 1×10 mm2 or 0.4×10 mm2
Stability ≤ 0.0005%  mains fluctuation
Goniometer Goniometer theta(θ)/theta(θ)
Diffraction circle semi-diameter 285mm
Scan range of θ -3° to +160°
Continuous scanning speed 0.006-96°/min
Setting speed of angle 1500°/min
Scan mode θ-θ or θ, θ; Continuous or step scanning
One way repeatability of θ ≤ 0.0002°
precision of θd or θs ≤0.005°
Minimal stepping angle 0.0001°
Record Unit Counter PC or SC
Maximal CPS 5x10^6 CPS
Proportion counter energy spectrum resolution ≤ 25%(PC), ≤ 50%(SC)
Detectable high voltage 1500-2100 continuous tune
High voltage of  the counter  differential or integral, automatic PHA, dead time emendation
ADX-DWZ System detector stability ≤ 0.01%
Micro Structure Micro Structure analysis, +/-0.5nm
Micro-Diffraction Micro sample or area, 2nm-19 um
Integrated performance Dispersion dosage ≤ 1μSv/h
Integrated stability of the system ≤ 0.5%
Dimension 1000 × 800 × 1640 mm